2.7 Setting the 0% and 100% Values
Procedure
Setting Values Using the 0% and 100% Keys
1.
With the source value and measurement value displayed, use the arrow keys to set the 0%
source value.
2.
Hold down the
0%
cursor key. The specified source value is assigned to the 0% value.
3.
Use the arrow keys to set the 100% source value.
4.
Hold down the
100%
cursor key. The specified source value is assigned to the 100% value.
When you hold down the 0% or 100% key,
this value is assigned to 0% or 100%.
When you display the percentage value,
you can check the 0% or 100% values.
Note
If you display the percentage on the screen using the DISPLAY key, you can view the assigned 0% and
100% values.
Setting Values Using the Setup Menu
1.
With the source value and measurement value displayed, press
SETUP
under Function 2.
2.
Use the arrow keys to select
Function2 Common Setup
, and then press
ENTER
.
3.
Use the arrow keys to select the
0% Value
value, and then press
ENTER
. The settings are
displayed at the bottom of the screen.
4.
Use the arrow keys to set the 0% value, and then press
ENTER
.
5.
Likewise, set the 100% value.
6.
Press the arrow key corresponding to
EXIT SETUP
. The settings are confirmed, and a screen
appears showing the source value and measurement value.
The cancel the settings, press
ESC
to return to the screen in step 2.
To initialize the settings, press the arrow key corresponding to
INIT SETUP
.
2-16
IM CA500-01EN