8
Connection Diagram (Loop power)
LOOP POWER
LOOP POWER
Black
Red
Digital multimeter
(DMM)
H (V)
L
DC voltage and
current generator
(Sink function)
Load
510 Ω/1 W
100 Ω
3.0000 V
H
L
-30.000 mA
Subdisplay
Calibration
point
Main display
Measured value
Calibration mode
5.2 Calibrating Output
Subdisplay
Fourth decimal place
Main display
Up to the third decimal place
Fourth decimal place
Second decimal place
Perform calibration according to Table 4.
) While holding down the SELECT and RANGE keys, turn the function switch from the
OFF position to the
position. (The “CAL” segment will be visible.)
2) Connect the CA0 to the reference device as shown in the diagram.
3) Start +5% calibration:
In the main display and the subdisplay, 0.0000 mA appears. (The display in the
subdisplay shows the value up to the fourth decimal place.)