IM CA11E-02E <Page 1>

■ Calibration the Generation Feature

Operation procedure:

<1> Warm up the standard.
<2> Before turning on the power of the CA11E calibrator, connect it to the standard.
<3> Turn on the power of the instrument.
<4> Simultaneously press and hold the 

 keys (shown in the figure in the 

 

“Assignment of Keys for Calibration” section below) for about 2 seconds to enter  
the calibration mode.

<5> Select the generation range to calibrate using the MEASURE/SOURCE selection  

switch and range selection rotary switch. The display unit shows “CAL,” “SOURCE,”  
“ON,” “0,” and the lower limit.

<6> Read the output value of the CA11E using the standard (digital multimeter), and using  

the 

 adjustment keys adjust the CA11E so that the output value is set to the offset  

value. Then press and hold the 

 input determination (ENTER) key for about 1 second  

to fix the setting. 

 

After fixing the setting, the display unit reading changes to “CAL,” “SOURCE,” “ON,” 

 

“FS,” and a full scale of the range.

<7> Read the output value of the CA11E using the standard (digital multimeter), and using  

the  

 adjustment keys adjust the CA11E so that the output value is set to the full  

scale value. Then press and hold the 

 input determination (ENTER) key for about  

1 second to fix the setting.

 

After fixing the setting, the display unit shows “0 FS” blinking. Re-press and hold 

 

the 

 input determination (ENTER) key for about 1 second to write the calibration  

coefficients to the EEPROM of the instrument. 

 

(This overwrites the previous calibration coefficients.) 

 

When this is complete, the instrument returns to the status in Step 5.

<8> Repeat Steps 5 to 7 for each range to be calibrated.

To return to the previous step:

<9> To return to the previous step without fixing the setting, press and hold the 

 input  

cancellation key for about 1 second. 
Note that this is possible only before writing to the EEPROM.

To return to the nomal operation mode:

<10> Simultaneously press and hold the  

 keys (shown in the figure in the 

 

“Assignment of Keys for Calibration” section below) for about 2 seconds, or press  
the [POWER] key to turn off the power once and then press it again to turn it back on.

■ Calibration Precautions

Connection for 20 mA SINK calibration
 

Connect the CA11E calibrator to the standard as show below:

CA11E

H

L

SINK

Lo Hi

DMM

current range

Constant-voltage

power supply

(Vo: 5 to 28 V)

■ Assignment of Keys for Calibration

When the CA11E calibrator is in the calibration mode, keys are assigned as specified here.

1

2

3

4

CAL

ON

SOURCE

0

Input cancellation key

Input determination

(ENTER) key

Digit to adjusted by 

the adjustment keys 

Mode switching 

(calibration or nomal) 

(Press and hold two keys 

simultaneously for 2 seconds.) 

Adjustment keys

(six keys) 

(For about 1 second)

(For about 1 second)

■ Calibration the Measurement Feature

Operation procedure:

<1> Warm up the standard.
<2> Before turning on the power of the CA11E calibrator, connect it to the standard.
<3> Turn on the power of the instrument.
<4> Simultaneously press and hold the 

 keys (shown in the figure in the 

 

“Assignment of Keys for Calibration” section below) for about 2 seconds to enter  
the calibration mode.

<5> Select the measurement range to calibrate using the MEASURE/SOURCE  

selection switch and range selection rotary switch. 
“CAL” and “MEASURE” appear and “FS” blinks on the display unit. 
(If a value nearly equivalent to full scale has been input, a measured value and  
“FS” appear.)

<6> Set up the standard in order to input the full scale value to the instrument. 

Wait until the reading stabilizes, then press and hold the 

 input determination  

(ENTER) key for about 1 second to fix the setting.

<7> After fixing the setting, “0” and “FS” indications on the display unit start blinking. 

 

Re-press and hold the  

  input determination (ENTER) key for about 1 second  

to write the calibration coefficients to the EEPROM of the instrument. 

 

(This overwrites the previous calibration coefficients.)

 

When this is complete, the instrument returns to the status in Step 5.

<8> Repeat Steps 5 to 7 for each range to be calibrated.

To return to the previous step:

<9> To return to the previous step without fixing the setting, press and hold the 

 input  

cancellation key for about 1 second. 
Note that this is possible only before writing to the EEPROM.

To return to the nomal operation mode:

<10> Simultaneously press and hold the 

 keys (shown in the figure in the 

 

“Assignment of Keys for Calibration” section below) for about 2 seconds, or press  
the [POWER] key to turn off the power once and then press it again to turn it back on.

Safety Precautions

The following manual, including this manual, are provided as manuals for the CA11E. 
Please be sure to read all manuals. 
IM CA11E-03EN 

(Getting Started Guide: For Safety) 

IM CA11E-01E  (User's Manual - 1 - : Generation, Measurement, and Other features) 

IM CA11E-02E  (User's Manual - 2 - : Calibration procedure and Specification): This manual

8.  Calibration Procedure

■ Calibration Procedure

To maintain a high level of accuracy, it is recommended that the CA11E HANDY CAL 
 be calibrated

 

annually.  

The “Selecting the Standards” section below presents calibration methods using the  
recommended standards.

■ Selecting the Standards

<Source feature>

Items to be  

calibrated

Names of  

standards

Range

Measuring  

range

Accuracy

Remarks

DCV

Digital  
multimeter
(DMM)

100 mV
1000 mV
10 V
30 V

Max. 110 mV
Max. 1.1 V
Max. 11 V
Max. 33 V

±(0.005%+5 μV)
±(0.005%+20 μV)
±(0.005%+200 μV)

±(0.005%+2 mV)

Model 7561
(YOKOGAWA)  
or equivalent

20 mA

Digital  
multimeter

20 mA (4-20 mA)

Max. 24 mA

±(0.01%+0.8 μA)

20 mA  
SINK

Digital  
multimeter,  
standard DC  
voltage generator

20 mA SINK

0.01 to 24 mA 
5 to 28 V

---

<Measurement feature>

Items to be  

calibrated

Names of  

standards

Range

Generated  

value

Accuracy

Remarks

DCV

Standard DC  
voltage generator

100 mV
1000 mV
10 V (1-5 V)
30 V

100 mV
1 V
10 V
30 V

±(0.01%/100 mV)
±(0.01%/1 V)
±(0.0%/10 V)
±(0.02%/30 V)

Model 2552
(YOKOGAWA)  
or equivalent

20 mA

Standard DC  
current generator

20 mA (4-20 mA)

20 mA

±(0.02%/20 mA)

---

■ Environmental Conditions for Calibration

   Ambient temperature: 

23 ± 1°C

   Relative humidity: 

45 to 75% RH

   Warm-up:     

Warm-up time as specified for the standard

■ Calibration Points

• The calibration points are as specified in the following tables.

• It is possible to independently select the necessary range to be recalibrated.

• Always calibrate the zero (0) point and full scale (FS) as a pair for generation.

 <Source>      

 

 

<Measurement>

Calibration points

Standard Value *1

Calibration points

Standard Value *2

100 mV

0

0 mV

100 mV

FS

100 mV

FS

100 mV

1 V

FS

1 V

1 V

0

0 mV

10 V

FS

10 V

FS

1000 mV

30 V

FS

30 V

10 V

0

0 V

20 mA

FS

20 mA

FS

10 V

*2: 
Set the Value to the standard as  

specified in the table.

30 V

0

0 V

FS

30 V

20 mA

0

0 mA

FS

20 mA

20 mA 
SINK

0.1

0.1 mA

FS

20 mA

*1:
Make adjustments to CA11E according  
to the reading of the standard  

(CA11E output Value) as specified in 

 

the table.

User’s

Manual

CA11E

HANDY CAL

(Voltage/Current Calibrator)

 

- 2 -

IM CA11E-02E

4th Edition: Dec. 2014

YMI

Printed in Korea

Keep this manual in a safe place so that

you can refer to it when necessary.