< 9. Calibration >
9-10
IM 11M12D01-01E
Enter
Semi-auto cal.
Ch1
Ch2
Ch3
Ch4
Ch5
Ch6
Ch7
Ch8
F9-8E.ai
Select calibration channel
Enter
F9-9E.ai
Ch1
Semi-auto cal.
Check settings
Span gas conc: 2 1 . 0 0 %
Zero gas conc: 0 . 9 8 %
Cal. time: 2 min 0 0 s
Points: Span-Zero
Go to setup to change value.
Start calibration
Cancel calibration
Figure 9.8 Channel Select Display
Figure 9.9 Semi-Auto Calibration Start
F9-10E.ai
21.00%
Ch1
1min./div.
0.98%
21.0 %O2
CAL. TIME
Semi-auto cal.
Figure 9.10 Semi-Auto Calibration Trend Graph
Starting Semi-Auto Calibration by a Contact input
1) Confirm that on the [Input contacts] display, [Calibration start] is selected.
2) In Measurement mode (with the Basic display displayed) when the contact input is detected then
Calibration starts.
3) To abort calibration, touch the Reject key (door with arrow icon). If you abort Span or Zero
calibration then calibration time is skipped and stabilization (settling) time starts. After that
touch the reject key (door with arrow icon) again, stabilization (settling) time is skipped and
abort calibration completely. If you touch the Enter key during stabilization (settling) time, abort
calibration of corresponding channel and next channel calibration will be started.
9.3.3 Starting Automatic Calibration
No user action is required to start automatic calibration. Automatic calibration starts at the preset time
and is repeated at the preset interval.
NOTE
• Before starting Semi-Auto calibration or Auto calibration, operate the calibration gas solenoid
valves and adjust calibration gas flow to 600 ± 60 ml/min for each sensor.
• If you skip Hold time, the “output hold” for output stabilization is released immediately after
Calibration time ends. (However, in this case, if “output hold” under maintenance status is set to
“enable”, the “output value” is constant after Calibration time ends). If stabilization time is skipped
and if calibration gas still remains in the sensor, the gas affect the accuracy of the output.