< 9. Calibration >
9-4
IM 11M12D01-01E
9.1.4 Characteristic Data from a Sensor Measured During Calibration
During calibration, calibration data and sensor status data (listed below) are acquired.
However, if the calibration is not properly conducted (an error occurs in automatic or semi-automatic
calibration), these data are not collected in the current calibration.
These data can be observed by selecting the [Detailed data] display key from the basic panel display.
For an explanation and the operating procedures of individual data, consult Subsection 10.1.1,
“Detailed Display.”
(1) Record of span correction ratio
Recorded the past ten span correction ratios.
(2) Record of zero correction ratio
Recorded the past ten zero correction ratios.
(3) Response time
You can monitor the response time provided that a two-point calibration has been done in semi-
automatic or automatic calibration.
(4) Cell’s internal resistance
The cell’s internal resistance gradually increases as the cell (sensor) deteriorates. You can
monitor the values measured during the latest calibration. However, these values include the
cell’s internal resistance and other wiring connection resistance. So, the cell’s degrading cannot
be estimated from these values only.
When only a span calibration has been made, these values will not be measured, and previously
measured values will remain.
(5) Robustness of a cell
The robustness of a cell is an index for predicting the remaining life of a sensor and is expressed
in a number on four levels.
9.2 Calibration Setup
9.2.1 Mode
There are three calibration modes (for zero and span calibration): manual mode, semi-auto mode
(started from the display or started by contact input) using preset calibration time and stabilization
(settling) time, and autocalibration at preset intervals. Restrictions related to each mode are described
here.
- When [Manual Mode] is selected.
In this mode, only Manual calibration can be performed. (Contact input used to start semi-auto
calibration has no effect. Autocalibration does not start at autocalibration start time.)
- When [Semi-Auto Mode] is selected.
In this mode, manual calibration or semi-auto calibration can be performed.
(Autocalibration does not start at autocalibration start time.)
- When [Auto Mode] is selected.
Any calibration mode is valid.
CAUTION
Indication check setup mode of [Auto] cannot coexist with [Auto] Calibration setup mode. Therefore,
If the mode of Indication check setup is set to [Auto], you cannot set the mode of Calibration setup to
[Auto], and vice versa.