4-6
IM AQ7280-01EN
5 Point Markers
Press the
5 Point Markers
soft key to display the following screen.
Sets marker.
Press this soft key to select
the marker that you are
operating.
Marker information
Press to set all markers
together.
Cursor (C)
Marker f
Marker F
Marker E
Marker N
Marker n
Note
• Set the markers in the order shown above, starting with the measurement start point (the distance
reference).
• The measured loss value differs depending on the specified approximation method.
• Set the correct position for marker E. The splice loss changes greatly depending on the position of E.
Presetting Markers
This function can be used to register markers that you set from the marker menu (2 Point Markers,
4 Point Markers, 5 Point Markers, 6 Point Markers) in advance and apply them on waveforms with a
single touch operation. Up to two sets of markers can be registered. This function is convenient when
measuring the same event repeatedly for each core such as in a multi-core fiber measurement.
1.
From the top menu, select
OTDR
.
2.
Press the
Marker
soft key to display the marker menu.
3.
Press the
Preset
soft key to display the preset menu.
Preset A marker
Selects the preset action
(Register, Delete, SetMarker)
Preset B marker
Preset information
display screen