4-44
IM AQ6376-01EN
Setting External Trigger Mode (for SMPL TRIG)
5.
Press
SETUP
. The soft key menu for sweep condition settings appears.
6.
Press the
MORE
soft key. The MORE 2/2 soft key menu is displayed.
7.
Press the
PLS LIGHT MEASURE
soft key.
8.
Press the
EXT TRIGGER MODE
soft key. External trigger mode is set.
Setting Trigger Conditions (
for SMPL TRIG)
9.
Press the
TRIGGER SETTING
soft key. The trigger condition setting menu is
displayed.
10.
Press the
EDGE
soft key. RISE switches to FALL, or vice versa.
11.
Press the
DELAY
soft key.
12.
Enter a delay time and press
ENTER
.
Note
When AVERAGE TIMES is set, sweeping stops when a number of external signals equaling the
specified sampling points times the "average times" is input.
4.16 External Trigger Measurement