4-32
IM AQ6376-01EN
4.11 Averaging Times Setting
Procedure
1.
Press
SETUP
. The soft key menu for sweep condition settings appears.
2.
Press the
AVERAGE TIMES
soft key. The averaging times setting screen is
displayed.
3.
Enter a number of averaging times using the rotary knob, arrow keys, or numeric
key pad.
4.
Press
ENTER
.
The value specified by the AVERAGE TIMES
soft key is
displayed.
Note
Increasing the averaging times reduces the sweeping speed but the S/N improves.
Explanation
AVERAGE TIMES
This key is used to set the average times for each point.
The setting range is 1 to 999.
If you press the COARSE key you can use the rotary knob or arrow keys to change
the numerical value in 1-2-5 steps. The value changes in 1 step if you do not press the
COARSE key.