Configuring a Firewall
You can close the TCP port when the communication interface (ETHERNET), remote
monitoring, and directory sharing are not in use.
1.
Press
SYSTEM
.
2.
Press the
MORE
soft key repeatedly until the MORE 2/4 menu is displayed.
3.
Press the
NETWORK SETTING
soft key.
The Ethernet setup menu appears.
4.
Press the
FIREWALL
soft key.
Each time you press the soft key, the setting toggles between ON and OFF.
When this is set to ON, the firewall is enabled.
Clearing Measured Data, Analysis Conditions, Parameters, Etc.
The PRESET key clears all internal settings of the AQ6376 except for the remote
interface (ETHERNET, GP-IB, and RS232) settings.
1.
Press
PRESET
. A confirmation message for clearing the message and YES and
NO soft keys appear.
2.
Press the
YES
soft key. The measured data and parameter settings are cleared.
If you do not want to clear them, press the
NO
soft key. The previous soft key
menu will return.
Explanation
Automatic Offset
When the AUTO OFFSET is set to ON, the offset of the internal amplifier circuit is
adjusted at time intervals of approximately 10 minutes. (Default: ON)
If the AUTO OFFSET is OFF, the auto-offset adjustment operation is not performed.
When you switch from OFF to ON, the offset adjustment is executed immediately.
When switched to ON, the offset adjustment is performed during repeat sweeping when
sweeping reaches 100%. During a single sweep, the offset adjustment is performed
when the sweep is completed.
During offset adjustment, a message indicating that the offset is being adjusted is
displayed in the lower-left corner and in the center of the top part of the screen.
Note
If, during offset adjustment, you use key operations or remote commands to execute a sweep,
the sweep will start after the offset adjustment is completed.
8.6 Other Settings