2-6
IM AQ6376-01EN
For Laser Type Light Sources
5.
Press the
EXECUTE LASER
soft key. A screen for specifying the wavelength of
the external light source is displayed.
6.
Select the laser wavelength value using the rotary knob or the arrow keys. The
allowed wavelength range is 1500–3400 nm.
7.
Press
nm/ENTER
. The wavelength value is set.
8.
Press the
EXECUTE
soft key. Wavelength calibration is executed. The previous
screen is displayed again after the calibration process ends.
9.
Press the
CANCEL
soft key during wavelength calibration to cancel the
wavelength calibration process.
For Gas Cell Absorption Line Type Light Sources
5.
Press the
EXECUTE GAS CELL
soft key. A screen for specifying the wavelength
of the external light source is displayed.
6.
Select the gas cell absorption line wavelength value using the rotary knob or the
arrow keys. The allowed wavelength range is 1500–3400 nm.
7.
Press
nm/ENTER
. The wavelength value is set.
8.
Press the
EXECUTE
soft key. Wavelength calibration is executed. The previous
screen is displayed again after the calibration process ends.
9.
Press the
CANCEL
soft key during wavelength calibration to cancel the
wavelength calibration process.
2.2 Wavelength Calibration