6-26
IM AQ6376-01EN
6.9 Optical Filter Characteristics Measurement
(Except Wavenumber Mode)
Procedure
Optical filter characteristics can be measured from the measured waveform of the light
input to the optical filter from the light source, as well as from the measured waveform
light output from the optical filter. (Except Wavenumber Mode)
Filter Measurement (Single Channel)
You can analyze a waveform whose number of modes is 1.
Filter Peak Analysis
1.
Press
ANALYSIS
. The soft key menu for analyzing measured waveforms
appears.
2.
Press the
ANALYSIS 2
soft key. The analysis function selection menu is
displayed.
3.
Press the
FILTER-PK
soft key. Analysis is performed, and the results are
displayed in the data area.
When changing the analysis parameters
4.
Continuing on from step 3, press the
PARAMETER SETTING
soft key. The
FILTER-PK analysis parameter setting screen is displayed. If a setting screen has
multiple pages, press the
NEXT PAGE
soft key to display the next screen.
5.
Move the cursor with the arrow keys or soft keys, and enter a setting value with
the numeric key pad. To select a check box, align the cursor then press the
SELECT
soft key.
6.
Press the
CLOSE WINDOW
soft key. The FILTER-PK analysis parameter setting
screen closes, and the soft key menu returns to the previous stage.
7.
Press the
ANALYSIS EXECUTE
soft key. Analysis is performed according to the
changed parameters, and the results are displayed in a list.