Calibration Table
Editing the User Calibration table. (Wavelength)
1.
Press SYSTEM.
2.
Press the WL OFFSET TABLE soft key. The wavelength calibration table appears
along with a soft key menu for editing calibration values.
3.
Press arrow soft keys to move the cursor to the wavelength to be edited, then
press the WL EDIT soft key. A screen for entering numerical values is displayed.
4.
Enter a value using the rotary knob, arrow keys, or numeric keypad.
5.
Press ENTER.
Note
Normally, it is not necessary to use this function.
Editing the User Calibration table. (Level)
1.
Press SYSTEM.
2.
Press the LVL OFFSET TABLE soft key. The level calibration table appears along
with a soft key menu for editing calibration values.
3.
Press arrow soft keys to move the cursor to the wavelength to be edited, then
press the LEVEL EDIT soft key. A screen for entering numerical values is
displayed.
4.
Enter a value using the rotary knob, arrow keys, or numeric keypad.
5.
Press ENTER.
Note
Normally, it is not necessary to use this function.
Explanation
Instrument’s Internal Calibration Wavelength Value
The wavelength value displayed in the soft key menu is for in the air or in a vacuum
depending on the measurement wavelength setting on the instrument.
Measurement Wavelength Setting List
Wavelength in the air
Wavelength in a vacuum
441.565nm
441.690nm
487.986nm
488.123nm
632.816nm
632.991nm
1152.274nm
1152.589nm
3.6 Wavelength Calibration