5.7 Warning Display Function
No. Message
Cause of Warning
No.1 - 49: Message generation after the execution of functions
1
Unsuitable resolution
Data may not be extracted completely, because the resolution
setting is not appropriate for the span and the sampling number.
2
Unsuitable level scale
A level scale larger than 5 dB/DIV has been set in the range fixed
mode (SENS:NORM HOLD).If a level scale is set to 5 dB/DIV or
more in the range fixed mode, data from the top and bottom of the
screen may not be properly displayed.
3
Unsuitable REF Level
Although an attempt was made to set the peak level of a waveform
to the reference level, the nearest value within the range was
chosen because the peak level value was outside the setting range
of the reference level values.
4
Unsuitable marker value
Although an attempt was made to set a maker value to the
reference level, the nearest value within the range was chosen
because the marker value was outside the setting range of the
reference level values.
5
<Auto Analysis> turned off
Selection of the <AUTO ANALYSIS> key was canceled.
6
<Auto Search> turned off
Selection of the <AUTO SEARCH> key was canceled.
7
Resolution mismatch between
traces
When the calculation between traces was performed, resolutions of
traces were set differently.
9
Trace * state changed
The HOLD state was canceled because the state of traces under
HOLD was changed from FIX to another state.
10
<Auto Ref Level> turned off
Selection of the <AUTO REF LEVEL> key was canceled.
11
<Auto Sub Scale> turned off
Selection of the <AUTO SUB SCALE> key was canceled.
14
Unsuitable sampling point
Sample point setting for using the smoothing function is missing.
15
<Marker Setting> changed
The Marker type was changed to Normal Marker.
17
<Level Unit> changed
The <LEVEL UNIT> key setting was changed.
18
<Power/NBW> trace fixed
The <POWER/NBW> trace setting in trace settings was changed to
FIX mode.
19
Inappropriate analysis
Because analysis was performed on a POWER/NBW trace, a
correct analysis result may not be displayed.
20
Optical power too high
The input light intensity is too high, and the measured waveform
may be saturated.
21
<Pulse Light Meas> turned off
The pulse light measurement mode was turned off.
22
<Sens> changed
The sensitivity setting was changed.
23
<Avg> changed
The average setting was changed.
24
<Sens><Avg> changed
The sensitivity setting mode and average setting were changed.
25
<Sens><Pulse Light Meas>
changed
The sensitivity setting mode and pulse light measurement mode
were changed.
26
<Avg><Pulse Light Meas>
changed
The average setting and pulse light measurement mode were
changed.
27
<Wavelength in AIR/VACUUM>
changed
The Air/Vacuum mode setting was changed as a result of loading a
file.
28
Fitting source trace data loaded
A fitting source trace was loaded from a curve fit waveform file.
29
<HCDR> turned off
HCDR mode was set to off.
30
<HCDR><Sensitivity> changed
HCDR mode and sensitivity setting were changed.
31
<HCDR><Avg> changed
HCDR mode and average setting were changed.
32
<HCDR><Sens><Avg> changed HCDR mode and sensitivity and average settings were changed.
33
<Resolution> changed
The resolution setting was changed.
34
<Resolution><Sensitivity>
changed
The resolution and sensitivity settings were changed.
35
<Res><Pulse Light Meas>
changed
The resolution setting and pulse light measurement mode were
changed.
36
<Res><Sens><Pulse Light>
changed
The resolution and sensitivity settings and pulse light measurement
mode were changed.