5.8 Sampling Point/Interval Settings
Procedure
The following are three ways in which the number of samples can be entered.
• Setting the number of samples directly
• Setting by the sampling interval
• Automatically setting the optimum number of samples or the sampling interval
according to the sweep width (span) and resolution setting
1.
Press
SETUP
. The soft key menu for sweep condition settings appears.
2.
Press the
SAMPLING POINT
or
SAMPLING INTERVAL
soft key to set the
sampling point or sampling interval, respectively. To automatically set the sampling
points and interval according to the span and wavelength (frequency) resolution
settings, press
SAMPLING POINT AUTO
. The sampling points or interval setting
screen is displayed.
If you pressed
SAMPLING POINT AUTO
, the sampling points and interval are set
automatically.
3.
Enter a number of samples or interval using the rotary knob, arrow keys, or
numeric key pad.
4.
Press
ENTER
. The sampling points or interval are set.
Automatic setting
Sets sampling points
Sets sampling interval
Note
• The instrument displays
if the settings for span, the number of sampling points, and
the resolution are inappropriate.
• See section 5.7 for the corrective actions when
is displayed.