YOKOGAWA AQ2200-631 User Manual Download Page 1

User’s
Manual

Yokogawa Electric Corporation

Model 810518803
AQ2200-631
10Gbit/s Optical Receiver
Remote Commands

IM 810518803-17E
1st Edition

Summary of Contents for AQ2200-631

Page 1: ...User s Manual Y okogawa Electric Corporation Model 810518803 AQ2200 631 10Gbit s Optical Receiver Remote Commands IM 810518803 17E 1st Edition...

Page 2: ...10Gbit s band optical interface by combining optional units such as BERT module 1 light source module 2 and optical modulation module 3 of the AQ2200 series thoroughly read relevant manuals 1 AQ2200 6...

Page 3: ...This Chapter describes how to set up the GP IB and Ethernet interfaces of this unit 3 Common Commands This Chapter describes the commands common to all AQ2200 series modules 4 Unit Specific Commands T...

Page 4: ...ions 1 1 1 4 Command Format 1 2 Chapter 2 Preparations for Connection 2 1 Connecting through GP IB 2 1 2 2 Connecting through Ethernet 2 2 Chapter 3 Common Commands 3 1 Common Commands 3 1 Chapter 4 U...

Page 5: ...te Control Functions The following shows the specifications of the GP IB and Ethernet remote control functions Specifications of GP IB remote control function Item Specifications GP IB address 0 to 30...

Page 6: ...SYST sysTEM syst SYSTEm SYSTeM System sYSTEM syST etc Incorrect description SYSTe SYS SYSTemm syste etc 4 Items described in in the syntax and response of the command show parameters Parameter name i...

Page 7: ...Follow the steps below to set the GP IB address 1 Press the SYSTEM key to display the SYSTEM screen 2 With the S or T key move the cursor to GP IB Address and press the Edit or ENTER key 3 With the te...

Page 8: ...nd press the Edit or ENTER key to select it 4 With the ten key pad or the S or T key change the numeric value and press the OK or ENTER key 5 Repeat steps 3 and 4 to input all parameters IP Address Su...

Page 9: ...051 01 00 SLOT n OPC description Obtains the command process status response status Type integer Range 0 1 0 Process is in progress 1 Process is completed SLOT n OPTions description Obtains the module...

Page 10: ...sults response result Type integer Range 0 Correct Value other than 0 Faulty If the value is not 0 the bit value corresponding to the fault item becomes 1 Bit Item Name 0 Upgrade Memory error 1 ID Inf...

Page 11: ...status 01 The temperature exceeds the upper limit of the operating temperature This may cause an operational fault 2 LOS alarm 0 Correct 1 Faulty 3 OVERLOAD alarm 0 Correct 1 Faulty SENSe n THReshold...

Page 12: ...output status response state Type discrete Range OFF ON INPut n POWer description Obtains the average optical input power value response power Type decimal SENSe n OVER LEVel syntax SENSe n OVLD LEVe...

Page 13: ...LEVel description Obtains the set value of the LOS detection level response los Type decimal Range 19 0 to 2 0 0 1dBm Step INPut n WAVelength syntax INPut n WAVelength length description Sets a wavel...

Page 14: ...030 Message Contents Remedy Type Command Error Command error Check the command and send it again CME 1031 Message Contents Remedy Type Syntax Error Syntax error Check the command syntax and send the c...

Page 15: ...re again DDE 2032 Message Contents Remedy Type Invalid ID information memory Update Memory of the Firmware is faulty Hardware is faulty DDE 2033 Message Contents Remedy Type Temperature limit error Te...

Page 16: ...sing Timeout Error to occur If this occurs wait for enough time after the command has been sent and send the next command or send the next command after the command has been completed by the OPC comma...

Page 17: ...Se n THReshold DATA Sets a data threshold value level 4 1 SENSe n THReshold DATA Obtains the data threshold value level 4 1 OUTPut n STATe Sets the data output to ON or OFF 4 2 OUTPut n STATe Obtains...