Analyzing Waveforms and Events

Measuring the Distance and Loss between Two Points

1.4, “Analyzing Measured Data” in the User’s Manual

You can place two markers on the waveform on the data display screen and measure the distance and loss between the markers.

1.

  Tap 

OTDR

 on the MENU screen.

2.

  Tap the 

Marker

 soft key and then the 

2 Point Markers

 soft key.

Drag

3.

  Tap the screen near the location where you want to display a cursor.

4. 

 Drag to adjust the position of the cursor on the screen. You can also use the rotary knob.

5.

  Tap a soft key.

: Distance measurement 

start position

A marker is placed at the 

cursor position.

: Distance measurement 

end position

Displays the measured results of the distance and loss between the markers 

(distance at the top and loss at the bottom within the frame)

When measuring the loss, if 

there are events 

(connections) between the 

markers, use TPA for the 

marker approximation 

method (see page 45).

Cursor

Start 

position

End 

position

Summary of Contents for AQ1210A

Page 1: ...User s Manual IM AQ1210 02EN 1st Edition AQ1210A AQ1215A AQ1210E AQ1215E AQ1215F AQ1216F OTDR Multi Field Tester Getting Started Guide...

Page 2: ...and how to use them For instructions on how to view the manual see page 17 AQ1210A AQ1215A AQ1210E AQ1215E AQ1215F AQ1216F OTDR Multi Field Tester Getting Started Guide IM AQ1210 02EN This guide The g...

Page 3: ...rictly prohibited Trademarks Microsoft Windows MS DOS are registered trademarks or trademarks of Microsoft Corporation in the United States and or other countries Adobe Acrobat and PostScript are eith...

Page 4: ...you ordered For reference the model name suffix code and specifications of the products are listed below front panel MODEL Suffix Code1 Specifications AQ1210A 2 wavelengths 1310 1550 nm DR 37 35 dB A...

Page 5: ...to the OTDR ports and the OPM port prior to shipping 3 1 25 mm diameter ferrule type for the optical power meter 4 SC type for the optical power meter No Instrument number When contacting the dealer...

Page 6: ...easurement port PPM option fixing screws 2 pcs Universal adapter SC Angled PC 735482 ASC For the OTDR port fixing screws 2 pcs For the PON optical power measurement port PPM option fixing screws 2 pcs...

Page 7: ...nt such occurrences French Une manipulation ou une utilisation incorrectes risquent de blesser l utilisateur ou d endommager l instrument Ce symbole appara t sur l instrument pour indiquer l utilisate...

Page 8: ...requirements The following symbols are used on this instrument Warning handle with care Refer to the user s manual or service manual This symbol appears on dangerous locations on the instrument which...

Page 9: ...uch as on the car dashboard or by a window inside a vehicle under the blazing sun near a stove or other heat source or other hot locations Your clothing may be damaged or you may be injured if you com...

Page 10: ...ort PON OPM port 1310 nm 1490 nm Do not apply light that is 27 dBm or greater to the high power OPM port PON OPM port 1550 nm Doing so may damage the OPM port Do Not Operate in an Explosive Atmosphere...

Page 11: ...nn sous un soleil de plomb proximit d une gazini re ou de toute autre source de chaleur ou dans tout autre endroit chaud Tout contact avec l lectrolyte chapp en raison d une fuite ou d une explosion d...

Page 12: ...utiliser l instrument en pr sence de gaz ou de vapeurs inflammables Cela pourrait tre extr mement dangereux Ne pas retirer le capot ni d monter ou modifier l instrument Seul le personnel YOKOGAWA qua...

Page 13: ...sification and requirements In addition this instrument complies with 21 CFR 1040 10 and 1040 11 except for deviations pursuant to Laser Notice No 50 dated June 24 2007 Laser Class 1M Label Using an o...

Page 14: ...0 nm CW 50 mW Pulse 200 mW Pulse width 20 s Duty 3 0 9 m 11 5 OTDR port PORT2 1M or 1 1625 nm CW 50 mW Pulse 200 mW Pulse width 20 s Duty 3 0 9 m 11 5 VLS port 3 3R 650 nm CW 5 mW 9 m 11 5 AQ1215E OTD...

Page 15: ...nditions 3 Applicable to models with the VLS option Regulations and Sales in Each Country or Region Waste Electrical and Electronic Equipment This directive is valid only in the EU This product compli...

Page 16: ...se them discard them in accordance with domestic law concerning disposal Take a right action on waste batteries because the collection system in the EU on waste batteries are regulated For instruction...

Page 17: ...le 1 Turn on the instrument 2 Connect the type B editionMicro B USB port on the top panel of the instrument to the type A USB port of the PC with the supplied USB cable For a description of the top pa...

Page 18: ...ow Keys 38 Using the Touch Panel 39 Entering Text 40 Setting the Language and Date and Time 41 Selecting the Language to Display 41 Selecting the Date and Time to Display 42 Setup 43 Setting Measureme...

Page 19: ...Emulation Software 61 Maintenance and Inspection 62 Replacing the Optical Adapter 62 Removing the Battery Pack 64 Recommended Part Replacement 66 Disposing of the Instrument 66 Specifications 67 Opti...

Page 20: ...al pulses REAL TIME key Press to start or stop the real time measurement of optical pulses ENTER key Press to confirm procedures and settings Arrow keys Press to change values move between digits move...

Page 21: ...rom a PC Note that the charging function and USB mass storage function cannot be used simultaneously Rear panel OTDR port PORT1 Used for OTDR measurement 1310 nm 1550 nm power checker measurement PC o...

Page 22: ...IM AQ1210 02EN 21 Rear and Side Panels Strap slot Stand Pull the stand out to use the instrument in a tilted position Protector Battery cover Component Names and Functions...

Page 23: ...SB cable and keep them away from heat sources General Handling Precautions Do Not Place Objects on Top of the Instrument Never place objects such as other instruments or objects that contain water on...

Page 24: ...cover 1 2 3 4 Strap loops Buckle Attaching the Shoulder Strap Strap loop Buckle Pass the belt back through the buckle and tighten it Attach the strap to the loops on both the upper left and upper righ...

Page 25: ...recautions that are included in the battery pack s user s manual French AVERTISSEMENT Ne pas installer ni d poser le pack de batteries lorsque l lectricit est aliment e par l adaptateur USB CA titre p...

Page 26: ...le hook Plug Battery pack Cable hook Power supply lead wires 5 Insert the battery pack so that the power supply lead wires are facing the bottom panel side from the lower right of the battery pack Mak...

Page 27: ...sure that the hooks on the battery cover enter into their holes on the case Battery cover Screws 7 Check that the battery cover is not loose and then fasten the attachment screws with a Phillips scre...

Page 28: ...that the rated supply voltage matches the voltage of the power supply Use only the supplied USB cable French AVERTISSEMENT Utiliser l adaptateur USB CA recommand par YOKOGAWA Utiliser l adaptateur US...

Page 29: ...1210E AQ1215E AQ1215F AQ1216F Supplied TypeC USB cable Note If the instrument s connector cover comes off bend the cover axle and reattach it For details on the USB AC adapter contact your nearest YOK...

Page 30: ...wer supply icon Charging Battery level indicator The battery pack may be too cold or too hot The battery pack disconnected Charging takes about 5 hours with the power turned off If the instrument is c...

Page 31: ...ace the power that has been lost through self discharge recharge the battery using the instrument once every six months Avoid storing the battery pack for an extended period of time when it is fully c...

Page 32: ...ector or optical port may be damaged If you use optical connectors that do not meet the specifications the instrument optical ports may be damaged Use optical connectors that are approved or used by n...

Page 33: ...nnecteurs optiques conformes aux sp cifications d faut de quoi les ports optiques de cet instrument pourraient tre endommag s Utiliser des connecteurs optiques homologu s ou utilis s par les entrepris...

Page 34: ...move the cable 4 Repeat steps 1 to 3 1 Open the optical port cover on the instruments top panel 2 Properly align the optical fiber cable s connector with the optical port and insert the connector You...

Page 35: ...wer switch turned on you can temporarily turn the power off to a sleep state The operation varies depending on how long you hold down the power switch Turning the Power Off If you hold down the power...

Page 36: ...ti core optical fiber cable measurement see sec 8 2 VLS Visible light source see sec 7 2 Advanced Analysis see sec 8 5 PPM option Light Source Light source see sec 7 1 Fiber Inspec Probe Fiber end fac...

Page 37: ...ply icon display Power supply icon blinks if the power supply ability to TypeC USB port is insufficient The battery pack may be dead if you continue use this instrument in this state Event waveform di...

Page 38: ...ed measurement in progress indication Auto save indicator see sec 2 1 Event Icon Display Manage data see chap 9 Utility button Shows the light source optical power meter and other screens see chap 7 S...

Page 39: ...ape Press Enter to display following setup item Items set from a list of options Move the cursor to the item that you want to select using the arrow keys or the rotary knob Press ENTER to confirm the...

Page 40: ...menu and so on Drag Drag refers to the act of pressing your finger against the screen and sliding your finger Pinch Out and Pinch In Pinch out refers to the act of pressing two fingers against the scr...

Page 41: ...splayed Entering Alphanumeric Characters Cursor Moves the cursor to the left Moves the cursor to the right Deletes the entered text Switches between uppercase and lowercase Switches the keyboard langu...

Page 42: ...Language to Display 1 Press MENU to display the MENU screen 2 Press SETUP to display the System Setup screen 3 Tap the Language button to display the language setup menu Language setup menu The langu...

Page 43: ...nute and second Confirms the settings The set date and time are displayed in the upper left of the screen System Setup Screen Set the date and time display format Off Year Month Day Time Day Month Yea...

Page 44: ...2 11 2 12 2 13 2 14 2 15 2 16 2 17 2 18 2 19 2 20 When the unit is duration AUTO 5sec 10sec 20sec 30sec 1min 3min 5min 10min 20min 30min Auto event detection on off Fiber in use alarm on off Turns co...

Page 45: ...001dB to 99 999dB An event bending is detected when this value is exceeded Setting a Launch Fiber Set the conditions for when a launch fiber is attached to the start or end face Set the approximation...

Page 46: ...h an alarm or the like Turns approximated line display on or off Displays an approximated line for waveform events Select the measurement distance unit This menu is not available on HJ suffix code Cal...

Page 47: ...to be measured 3 Press REALTIME to start a measurement Pressing it again stops the measurement Data display screen Waveform is updated during measurement Laser on indication Real time measurement in...

Page 48: ...configuring this feature start real time measurement according to the steps on the previous page 4 Move the marker to the position where you want to monitor the waveform in detail 5 Tap the Refresh R...

Page 49: ...a measurement If you press AVG again before measurement completion the measurement stops at that point Data display screen Displays the averaged waveform Laser on indication Averaged measurement in pr...

Page 50: ...screen near the location where you want to display a cursor 4 Drag to adjust the position of the cursor on the screen You can also use the rotary knob 5 Tap a soft key Distance measurement start posit...

Page 51: ...oint Markers soft key Drag 3 Tap the screen near the location where you want to display a cursor 4 Drag to adjust the position of the cursor on the screen You can also use the rotary knob 5 Tap a soft...

Page 52: ...creen near the location where you want to display a cursor 4 Drag to adjust the position of the cursor on the screen You can also use the rotary knob 5 Tap a soft key Measurement start position A mark...

Page 53: ...display a cursor A cursor appears 4 Drag to adjust the position of the cursor on the screen You can also use the rotary knob Drag Resets waveform zoom to the original size Zooms in horizontally Zooms...

Page 54: ...e an event analysis Measurement reference point S Event number fault point connection etc Event detected as end of fiber Event analysis results event information The selected event information is disp...

Page 55: ...d Switches to the waveform view Dist between events Distance from the measurement reference point Change the event type Reflection Positive splice loss Negative splice loss Bending loss macro bending...

Page 56: ...2 Display a waveform on the screen 3 Tap the FILE icon and then REPORT to display the file list screen File name Exports a report file File list screen The waveform shown on the data display screen is...

Page 57: ...and then FILE to display the file list screen 3 Select the waveform data file to export to a report file For the procedure to select multiple files see section 9 6 in the User s Manual IM AQ1210 02EN...

Page 58: ...icon A report file will be created in the same folder as selected file Export report icon Report file Label information Event Icon Display Waveform display Event search conditions Event analysis resul...

Page 59: ...sections generate loss The measured result is displayed as a waveform that has distance represented in the horizontal direction and loss level represented in the vertical direction On the waveform de...

Page 60: ...the section from the measurement start point to the open end are displayed in order from the start point Incident ray Backscatter Splice Connector Bend Open end Optical fiber cable Splitter Splice los...

Page 61: ...ctor Using a connector to connect two optical fibers is different from splicing them together in that a small gap remains between the two fibers Because this gap has a different index of refraction re...

Page 62: ...tware This software also has a PDF report creation feature which is convenient for creating construction reports Waveform data saved in SOR format using the instrument can be displayed on the PC using...

Page 63: ...the emitted light to accidentally enter the eye This can cause eye damage or vision impairment French AVERTISSEMENT Lors du nettoyage de l metteur de lumi re mettre l instrument hors tension pour emp...

Page 64: ...gth 4 mm OTDR port Light source port cover OPM port OTDR port cover Optical adapter Guide hole Installation 1 Check that the instrument is turned off 2 Open the OTDR ports or the OPM port SPM HPM or P...

Page 65: ...ling precautions that are included in the battery pack s user s manual French AVERTISSEMENT Ne pas installer ni d poser le pack de batteries lorsque l lectricit est aliment e par l adaptateur USB CA t...

Page 66: ...attery cover toward you lift and remove Battery cover 3 Insert your finger into the hole so that the battery pack can be removed Battery pack Plug 4 Lift the battery pack with your finger 5 Disconnect...

Page 67: ...tervals listed below Part Name Service Life Notes Backup battery lithium battery 5 years Send to factory to replace LCD Approx 50 000 hours Send to factory to replace Measurement connector ferrule One...

Page 68: ...y 0 05 dB dB 0 03 dB dB 0 05 dB dB 0 03 dB dB Sampling resolution 5 cm min 2 cm min 5 cm min 2 cm min Number of sampling points 256000 max 256000 max 256000 max 256000 max Distance range 0 1 km to 256...

Page 69: ...512 km 0 1 km to 512 km Pulse width 3 ns to 200 s 3 ns to 20 s Return loss measurement accuracy 2 dB 2 dB Distance measurement accuracy 1 m measured distance 2 10 5 1 sampling resolution Compliant fi...

Page 70: ...50 nm Noise level 0 5 nW 1310 nm 50 nW 1310 nm Uncertainty at reference conditions1 5 5 0 5 dB 1 CW 1310 2 nm Standard High Power PON at 1310 nm 1550 2 nm PON at 1550 nm spectral width 10 nm or less i...

Page 71: ...ns Multi Core Fiber Measurement Installation Completion Notification Smart Mapper Auto Loss Test Report Creation Display section Display 5 7 inch size color TFT LDC1 Display pixels 640 horizontal 480...

Page 72: ...protrusions Weight Approx 1 kg including the battery pack Safety standards Compliant standard EN 61010 1 IEC 61010 1 EN 60825 1 2014 laser safety IEC 60825 1 2007 laser safety FDA 21 CFR 1040 10 and 1...

Page 73: ...AQ1210 02EN External Dimensions 3 7 210 3 7 8 27 0 15 0 15 69 2 72 148 5 83 If not specified the tolerance is 3 However in cases of less than 10 mm the tolerance is 0 3 mm Unit mm approx inch Specific...