10-1
IM AQ1200-01EN
Chapter 10 PON Power Meter (Option)
PON Power Meter (Option)
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
App
Index
20
10.1 Making Preparations for Measurements
Remove the optical fiber cables from the Instrument and close the optical connector covers, or make
sure that the power meter is not receiving any light, and then start the PON power meter zero set
procedure.
Procedure
Performing Zero Set
1.
Using the rotary
knob
and
ENTER
, select
PON_OPM
(PON power meter).
2.
Press
SETUP
and then the
OPM Setup
soft key to display the following screen.
Performs zero set
Explanation
Perform zero set whenever necessary, such as after you have turned on the power or when the
ambient temperature changes. Performing zero set adjusts the internal deviation of the optical power
measurement section and enables you to obtain more accurate absolute optical power values.