5. BASIC OPERATING PROCEDURES
IM 1E7B0-02E
5-3
5.2.1
Flow Rate Data Display Mode
•
The flow rate data display mode indicates instantaneous flow and totalized flow
values.
The ADMAG AE can display 12 types of flow rate data.
•
This function can be set in the parameter "d1" of the flow converter.
•
For changing from setting mode to flow rate data display mode, press "SET" key
while pressing "SHIFT" key.
•
When a BT is used, call up the "D01 DISP SELECT" parameter to select functions.
Display item
Content
ADMAG AE Setting
BT Setting
Forward
Flow Rate %
Instantaneous flow rate is displayed within a
range of -8(or -108) to 108 for the span.
Actual forward
Flow Rate
The actual meter rate of instantaneous flow rate
is displayed.(See note 1.)
The decimal place is the same as for the span
setting. However,since a decimal point set at the
most sifnigficant bit cannot be displayed.
It is automatically shifted 1 digit to the right. (BT setting
of 0.0001 is displayed as 0.000 on the ADMAG.)
Forward flowrate
totalization values
Reverse flowrate
totalization values
Display forward flow rate totalization value
Display reverse flow rate totalization value
Differential
Between the
forward and
reverse
totalization
values
Alternate display
of forward
flowrate % and
forward
flowrate
totalization
values
Differential totalization, between
Forward-totalization and reverse totalization,
is displayed.
Display alternately between display of "RATE
(%)" and "FOR. TTL" every 4 second interval.
D01 : DISP SELECT
RATE(%)
D01 : DISP SELECT
DIF. TOTAL
D01 : DISP SELECT
RATE
D01 : DISP SELECT
FOR.TOTAL
D01 : DISP SELECT
REV.TOTAL
D01 : DISP SELECT
RATE%/FOR.TTL
Note 1 :
The LCD can display the following combination of units (by selecting a parameter)
Units other than those shown below can be displayed by attaching the provided unit labels .
l (liters) /h, l (liters) /m, m
3
/h, m
3
/m, gal/h, gal/m