7. OPERATION VIA BRAIN TERMINAL
IM 1E7B0-02E
7-14
7.2.19 User-Defined Units Via the BT200
This function displays the instantaneous flow rate indicated in "A20 FLOW RATE" in
units other than those selectable with B04. If the specific gravity of the fluid is known,
the instantaneous flow rate can be displayed in weight units.
(1) User-Defined Units for Instantaneous Flow Rate
Example :
Displaying the flow rate of a fluid (its specific gravity is 2) in weight (kg)
in a flow rate span of 10m
3
/h. When the flow rate is 100%, 20,000kg is
displayed.
Display
Description
PARAM
D02:FL USER SEL
PROVIDED
D03:FL USER SPAN
20000.0
D10:FL USER UNIT
kg/h
DATA
ESC
DIAG
DATA
ESC
DIAG
PARAM
A10:FLOW RATE(%)
100.0%
A20:FLOW RATE
20000.0 kg/h
A30:TOTAL
12345
DATA
ESC
DIAG
DATA
ESC
DIAG
Select "PROVIDED" in "D02."
For item "D03," set the value to be displayed in "A20 FLOW RATE"
when the flow rate is 100% in the span set in "B03FLOW SPAN." Set
the unit in "D10".
Set the user-defined unit in "D10."
The instantaneous flow rate in user-defined unit can be displayed in
"A20 : FLOWRATE."
(2) User-Defined Units for Totalization Values
User-defined unit can be added to the totalization display in "A30 TOTAL".
Display
Description
Set the user-defined unit in the item "E10."
The totalization value in user-defined unit can be displayed in "A30 TOTAL" .
PARAM
E04:TOTAL SET
INHIBIT
E05:TL SET VALUE
0
E10:TL USER UNIT
ESC
DIAG
DIAG
DATA
PARAM
A10:FLOW RATE(%)
100.0%
A20:FLOW RATE
40000.0 kg/h
A30:TOTAL
12345
ESC
CLR
DATA