8. ACTUAL OPERATON
8.2.3
Error Description and Countermeasure
Display detail
BRAIN Terminal
HART Communicator
Indicator
of SE
Description
Countermeasure
---
GOOD
---
ERROR
Err-02 uP FAULT
Err-03 EEPROM FAULT
Err-05 A/D FAULT
Err-06 SIGNAL
OVERFLOW
Err-07 COIL OPEN
Err-08 SPAN VEL.>10m/s
Err-09 SPAN VEL.<0.3m/s
Err-10 P.SPAN>1000p/s
Err-11 P.SPAN>500p/s
Err-12 P.SPAN>25p/s
Err-13 P.SPAN>15p/s
Err-14 P.SPAN>10p/s
Err-15 P.SPAN>5p/s
Err-16 P.SPAN<.0001p/s
Err-17 T.SPAN>1000p/s
Err-18 T.SPAN>.0001p/s
Err-19 4-20 LMT ERROR
Err-20 EMPTY PIPE
Err-21 MULTIRANGE
ERROR
Err-22 COIL SHORT
Err-23 EEPROM NO SET
Err-24 EEPROM
DEFAULT
Err-25 ---
Self test OK
Error
uP fault
EEPROM fault
A/D fault
Signal overflow
Coil open
Vel. span>10m/s
Vel. span<0.3m/s
P.span>1000p/s
P.span>500p/s
P.span>25p/s
P.span>15p/s
P.span>10p/s
P.span>5p/s
P.span<.0001p/s
T.span>1000p/s
T.span<.0001p/s
Analog lmt error
Empty pipe
Multi range err
Excitation failure
EEPROM no set
EEPROM default
Dev id not
entered
Normal status
Alarm occurrence
(At occurrence of error
No.2 to 25)
Microprocessor failure
EEPROM failure
A/D converter failure
Excessive input signal
Flow tube coil open-circuit
Span velocity setting
exceeds 11m/s.
Span velocity setting is
0.2m/s or less.
Pulse output rate exceeds
1100p/s at 50% duty.
It exceeds 1000p/s at
0.5ms pulse width.
Pulse output rate exceeds
500p/s at 1ms pulse width.
Pulse output rate exceeds
25p/s at 20ms pulse width.
Pulse output rate exceeds
15p/s at 33ms pulse width.
Pulse output rate exceeds
10p/s at 50ms pulse width.
Pulse output rate exceeds
5p/s at 100ms pulse width
Pulse output rate is
0.00005p/s or less.
Totalization rate exceeds
1100p/s.
Totalization rate is
0.00005p/s or less.
Analog low limit setting
exceeds analog high limit
setting.
Pipe is not filled with
fluid or insulating material
attached to electrodes.
“Span” setting exceeds
“Forward span2” setting
at Auto 2 ranges.
Excitation circuit failure
Default values have not
been set in EEPROM.
EEPROM has been
initialized.
Device ID has not been
entered.
Status
group 1
Status
group 2
Status
group 3
—
—
—
Contact the nearest
Yokogawa office or
service center.
Check the followings:
• Mix of the signal, power
supply, excitation and
other cables
• Stray currents in the
measured fluid
• Incorrect grounding
Contact the nearest
Yokogawa office or
service center.
Change settings.
Fill the flow tube with
fluid or clean the flow
tube inside.
Change settings.
Contact the nearest
Yokogawa office or
service center.