8. ACTUAL OPERATON
8.2.2
Self-diagnostics Using HART Communicator
Self-diagnostics of the ADMAG SE and check of incorrect data setting can be carried
out with the HART Communicator. There are two methods for self-diagnostics of the
ADMAG SE, self-diagnostics for every transmission and manually executing the “Self
test” command. When an error message appears, follow “8.2.3 Error Description and
Countermeasure.”
1. Device setup
2. Diag/Service
1. Test device
1. Test/Status
ADMAG SE :
Test/Status
1 Status
2 Self test
HELP
SAVE
HOME
ENTER
1
2
Select “Self test”.
If there is no error detected, “Self test OK” will be displayed.
Press OK (F4).
When an error occurs, an error message appears and the results
of self-diagnostics appear in the “Status”.
Call up “Status”.
The status menu is separated 3 groups. About items of each
group, see “8.2.3 Error Description and Countermeasure”.
Select the desired group.
If there is no error, the result of diagnostics is indicated as
“OFF”. If “ON” is indicated, a countermeasure for that error is
necessary.