IM 1E10C1-E
5-17
5. FUNCTION AND DATA SETTINGS
5.5.12 Alarm Output
• This function is for status output from PLS/ALM+, PLS/ALM– terminals, when an
alarm occurs.
PLS/ALM+, PLS/ALM–terminals are for common use with pulse, alarm and other
status output functions.
Therefore, in case this function is used, other functions are not available to use.
• All of the alarms are able to output except empty pipe detection function that can be
selected in parameter No. “n07” as out of selection.
• The status goes from close to open (OFF) during alarming.
The alarm output can be selected “01” in parameter No.
“b10” and P+, P– terminals are only for alarm output.
The empty pipe output selection can be set in parameter
No. “n07”.
Display
Description
Code
00
01
02
03
04
05
Pulse output
Alarm output
Forward / reverse flow measurement
Automatic two range switching
Alarm output at low flow limit
Totalization switch
Content
Content
Code
00
01
ALARM
NO ALARM
Default: 01
Default: 00
* Item “n” can be called up in setting “55” in parameter No. “L02”.
5.5.13 Data Settings Enable / Inhibit
• This function can inhibit to change all data except parameter No. “ L01”.
However, auto zero adjustment function can work, if it has been set in parameter No.
“C01”.
And the preset totalization value function also can work, if it has been set in param-
eter No. “E04”.
The data settings inhibit item can be set “00” in
parameter No. “L01”.
Displayed
Description
Default: 01
Code
00
01
INHIBIT
ENABLE
Content
NOTE