13-5
IM 765501-01E
GP-IB
Communications
3
2
1
6
5
4
9
8
7
12
11
10
15
14
13
13
18
17
16
Index
App
13.4 Setting the GP-IB Address
Procedure
MEASURE
STORE
AUTO
RANGE
MENU
MATH
NULL
COMPARE
MEASURE
VALUE
LOCAL
POWER
MISC
TIME
KEY LOCK
SHIFT
V I
MODE
AUTO
RANGE
LIMIT
VS IS
MENU
SOURCE
OUTPUT CONTROL
SWEEP
START
ZERO
SOURCE
TRIG
MODE
OUTPUT
RANGE
Ω
RECALL
DISPLAY
ESC
• To exit the menu during operation, press
ESC
at the upper left of the operation
panel.
• In the procedural explanation below, the
phrase “
rotary knob, numeric keys, <,
and >
” are used to refer to the operation
of selecting or setting items or entering
values using the
rotary knob
,
numeric
keys
, , and keys (BS key and
right arrow key). For details on the
operation using the rotary knob, numeric
keys, , and , see sections 3.8 or
3.9.
1.
Press
MISC
to display the MISC menu.
2.
Press the
Remote I/F
soft key to display the communication menu.
Setting the GP-IB Address
3.
Press the
GPIB
soft key.
4.
Use the
rotary knob
or
numeric keys and < and > keys
to set the GP-IB
address.
5.
If you use the numeric keys, press the
Enter
soft key to confirm the setting.
Explanation
GP-IB Address
Each device that can be connected via GP-IB has a unique address within the GP-IB
system. This address is used to distinguish the device from others. If you are
connecting the GS610 to a controller such as a PC, select the GP-IB address of the
GS610 on the PC side.
Selectable range: 0 to 30