8.2
NULL Computation
Procedure
MEASURE
STORE
AUTO
RANGE
MENU
MATH
NULL
COMPARE
MEASURE
VALUE
LOCAL
POWER
MISC
TIME
KEY LOCK
SHIFT
V I
MODE
AUTO
RANGE
LIMIT
VS IS
MENU
SOURCE
OUTPUT CONTROL
SWEEP
START
ZERO
SOURCE
TRIG
MODE
OUTPUT
RANGE
Ω
RECALL
DISPLAY
ESC
• To exit the menu during operation, press
ESC
at the upper left of the operation
panel.
• In the procedural explanation below, the
phrase “
rotary knob, numeric keys, <,
and >
” are used to refer to the operation
of selecting or setting items or entering
values using the
rotary knob
,
numeric
keys
, , and keys (BS key and
right arrow key). For details on the
operation using the rotary knob, numeric
keys, , and , see sections 3.8 or
3.9.
Turning NULL Computation ON/OFF
1.
Press
NULL
. The NULL key illuminates, and the NULL computation is enabled.
2.
If you press
NULL
again, the NULL computation is disabled, and the NULL key
turns OFF.
Setting the NULL Value
1.
Press
SHIFT+NULL
. The retrieved NULL value is displayed.
2.
Use the
rotary knob
or
numeric keys and < and > keys
to set the NULL value.
3.
If you use the numeric keys, press a soft key corresponding to the desired unit
or the
Enter
soft key to confirm the setting.
Explanation
NULL Computation
The measured value when NULL computation switches from OFF to ON when you press
the NULL key is used as a NULL value. While NULL computation is ON, the value
obtained by subtracting the NULL value from the measured value is displayed.
NULL Value
The retrieved NULL value applies to all measurement range settings.
<<Corresponding Command Mnemonic>>
:CALCulate:NULL[:STATe]
:CALCulate:NULL:OFFSet