6-7
IM 765501-01E
Sweep
3
2
1
6
5
4
9
8
7
12
11
10
15
14
13
18
17
16
Index
App
6.3
Selecting the Termination Mode
Procedure
MEASURE
STORE
AUTO
RANGE
MENU
MATH
NULL
COMPARE
MEASURE
VALUE
LOCAL
POWER
MISC
TIME
KEY LOCK
SHIFT
V I
MODE
AUTO
RANGE
LIMIT
VS IS
MENU
SOURCE
OUTPUT CONTROL
SWEEP
START
ZERO
SOURCE
TRIG
MODE
OUTPUT
RANGE
Ω
RECALL
DISPLAY
ESC
• To exit the menu during operation, press
ESC
at the upper left of the operation
panel.
• In the procedural explanation below, the
phrase “
rotary knob, numeric keys, <,
and >
” are used to refer to the operation
of selecting or setting items or entering
values using the
rotary knob
,
numeric
keys
, , and keys (BS key and
right arrow key). For details on the
operation using the rotary knob, numeric
keys, , and , see sections 3.8 or
3.9.
1.
Press
SWEEP
to display the SWEEP menu.
2.
Press the
Last
soft key to select the sweep termination mode.
Explanation
Sweep Termination Mode
Return (Rtn)
The source level after the sweep operation is complete is reset to the level before the
sweep operation was started.
In return (Rtn) mode, a trigger for resetting the source value is needed in addition to the
trigger for the measurement. Therefore, note that one additional trigger is needed for the
GS610 to enter the start wait condition.
Sd
Trigger
Meas
Trigger Trigger Trigger Trigger
Sd Meas Sd Meas Sd Meas Sd Meas
Trigger
Sd
Source measure
cycle
Source level
Stop value
Start value
Source level
before sweep
Waiting for
start
Sweep state
Start
Waiting for
start
Sweeping
• Sweep operation when the termination mode (Last) is set to return (Rtn)
Sd: Source delay
Meas: Measurement delay + measu computation