Setting
the
Measurement
Conditions
4-7
IM 755601-01E
4
4.4 Using the Contact Check Function
Keys
NORMAL
MEAS&
LOCK
SAMPLE
EXTERNAL
REFERENCE
MANUAL
FAST
HI
HI LIMIT
EXIT
SETUP/MEAS
CHECK
MISC
I/F
MEAS
TIME
STORE
DATA
RECALL
DATA
LOCAL
ENTER
LIMIT
MODE
TRIG
MODE
SHIFT
R/%
REF
LIMIT
TRIG
7
1
0
.
2
3
5
4
6
8
9
LO LIMIT
IN
LO
M
Ω
M
Ω
k
Ω
Ω
,%
k
Ω
Ω
Ω
%
%
HIGH
SPEED
STORE
DATA
RECALL
DATA
CHECK
REMOTE
Indicator is ON while contact check
is in progress
Procedure
The items that are specified or selected are confirmed when the ENTER (
Ω
,%) key is
pressed. When confirmed, the display returns to the top menu of the SETUP mode.
To exit from a menu in the middle of the operation, press the EXIT (R/%) key. The
display returns to the top menu of the SETUP mode.
Switching to the SETUP mode
1.
Press the
SHIFT
SETUP/MEAS
R/%
key to switch to the SETUP mode. If the “MEAS & LOCK”
indicator is off, this operation is not necessary.
Using the contact check
2.
4.
3.
6.
5. Numerical keys
(Contact check level)
CHECK
ENTER
SHIFT
REF
LIMIT
M
Ω
Ω
,%
ENTER
Ω
,%
ENTER
Ω
,%
[Main display]
If the value entered using the numerical keys is not correct, an error (810) is displayed
for approximately one second, and the display returns to step 5.
Explanation
Contact check function
If the result of the contact check is larger than the specified check level, an error is
generated. The time duration of the check is 2 ms, and the check current is 50 mA.
The timing to perform the contact check can be selected from the following choices.
OFF:
Contact check is not performed.
bEF:
Contact check is performed before the measurement. If an error is detected, “–nC–” is
displayed and the “HI” indicator ( ) turns ON. “HI” “NO CONTACT” signal is output
from the handler interface.
AFt:
Contact check is performed after the measurement. If an error is detected, “–nC–”
is displayed and the “HI” indicator ( ) turns ON. “HI” “NO CONTACT” signal is
output from the handler interface.
Setting range of the check level
1 to 30
Ω
(1
Ω
resolution).