11-12
IM 755601-01E
11.5 Replacing the Power Fuse
WARNING
• To prevent the possibility of fire, use only a fuse having the specified rating
(voltage, current, and type).
• Make sure to turn OFF the instrument and unplug the power cord before
replacing the fuse.
• Never short the fuse holder.
Specified Rating
The power fuse used on this instrument is specified as follows.
Maximum rated voltage: 250 V
Maximum rated current: 315 mA (100/120 V), 160 mA (220/240 V)
Type:
Time lag
Standard:
VDE/SEMKO/UL/CSA/SEV certified
Part number:
A1345EF (100/120 V), A1342EF (220/240 V)
Replacement Procedures
Follow the procedures below to replace the power fuse.
1.
Turn OFF the power switch.
2.
Unplug the power cord from the power outlet.
3.
Press the fuse holder section to the right of the power connector with a Philips
screwdriver, for example, and turn it to the left in order to remove the fuse holder.
Place a new fuse in the fuse holder, and place the fuse holder back in its original
position.