11-10
IM 755601-01E
Communication Commands Related to the Adjustment Method
The functions in the CALibrate group are valid only during the range adjustment mode. Read pages 10-1 to 10-4
carefully before executing commands in this group.
:CALibrate
:
ADJust
CANCel
RANGe
STATe
?
?
;
WRITe
ZERO
FULL
?
<Resistance>
:
MIN
DOWN
<Space>
<Resistance>
?
<Space>
OFF
ON
<NRf>
?
11.4 Adjustments
:CALibrate?
Function
Queries all settings related to adjustments.
Syntax
:CALibrate?
Example
:CALIBRATE?
→
:CALIBRATE:ADJUST:STATE 1;
RANGE 1.0E+05
:CALibrate:ADJust?
Function
Queries all setting related to range adjustment.
Syntax
:CALibrate:ADJust?
Example
:CALIBRATE:ADJUST?
→
:CALIBRATE:ADJUST:STATE 1;
RANGE 1.0E+05
Description The response is the same as that of the
“
:CALibrate?
” query.
:CALibrate:ADJust:CANCel
Function
Terminates the range adjustment without
writing the adjusted values to the EEPROM.
Syntax
:CALibrate:ADJust:CANCel
Example
:CALIBRATE:ADJUST:CANCEL
Description This command is equivalent to the
“
:CALibrate:ADJust[:STATe] OFF
” command.
:CALibrate:ADJust:FULL?
Function
Executes the full scale measurement for the
current range and determines the new gain
adjustment value. At the same time, queries
the value that is measured using the new
adjusted value.
Syntax
:CALibrate:ADJust:FULL? {<Resistance>}
<Resistance> = Resistance of the standard
resistor
0.00000
Ω
to 120.000 M
Ω
(The setting
resolution is 1/10 of the display
resolution)
Response
<NR3> form
Example
:CALIBRATE:ADJUST:FULL? 100.001KOHM
→
1.05 (Normal)
:CALIBRATE:ADJUST:FULL? 100.001KOHM
→
9.9E+37 (When an error occured)
Description • If the result of the full scale measurement of
the range is inappropriate, an error 872
occurs, and “9.9E+37” is returned as a
response. In addition, the adjustment does
not move to the next range.
• If the full scale measurement is completed
successfully, the adjustment automatically
moves to the next range. Therefore, if you
wish to make another full scale measurement
at the same range, issue the
“
:CALibrate:ADJust:RANGe DOWN
” command
to set the adjustment range back by one
before making the measurement again.
• If the zero measurement
“
:CALibrate:ADJust:ZERO
” has not been
executed successfully, this command cannot
be executed. Otherwise, an error 873 occurs.