Completing the adjustment in the middle of the range
5.
Press the ENTER key when the adjustment range is blinking on the main display.
“WritE” is displayed.
6.
When the ENTER key is pressed, the adjusted values existing in the range previous
to that which was blinking are written to the EEPROM and the adjustment
terminates. Reboot the instrument.
Canceling the range adjustment
5.
Press the EXIT(R/%) key when the adjustment range is blinking on the main
display. The adjustment is aborted and “AdJ” is displayed. In this case, the
adjusted values are not updated.
*1 “0.0000”
Ω
blinks for the 755601.
*2 “1.0000”
Ω
blinks for the 755601.
Explanation
Change in the value entry
While entering values, the <(REF) and >(LIMIT) keys become backspace and clear keys,
respectively.
Adjustment range
The adjustment ranges are as follows:
1
Ω
, 10
Ω
, 100
Ω
, 1 k
Ω
, 10 k
Ω
, 100 k
Ω
, 1 M
Ω
, 10 M
Ω
, and 100 M
Ω
11.4 Adjustments