10-64
IM 701730-01E
Explanation
This section explains the setup procedures for creating determination zones based on a
reference waveform and performing GO/NO-GO determination on whether the waveform
exits or enters the determination zones.
Determination Target Waveform
The waveform on which GO/NO-GO determination is performed is CH1 to CH4(2)*,
Math1, or Math2.
*
On the DL1735E/DL1740E/DL1740EL, the waveform is CH1 to CH4, Math1, or Math2. On
the DL1720E, you can select CH1, CH2, and Math1.
Determination Zones
• The reference waveform for creating the zone is selected from the displayed
waveforms, and six determination zones, Zone1 to Zone6, are created and registered.
The selectable range of the zone is as follows:
Selectable range in the up and down direction:
8 divisions above and below the reference waveform.
Selectable range in the left and right direction:
5 divisions to the left and right from the screen center.
• You can select the reference waveform for creating determination zones from CH1 to
CH4(2)*, Math1, and Math2.
*
On the DL1735E/DL1740E/DL1740EL, you can select from CH1 to CH4, Math1, and Math2.
On the DL1720E, you can select CH1, CH2, and Math1.
• The zone can be edited entirely or partially with respect to the reference waveform.
• Registered zones can be modified.
Determination Condition and Instrument Action When the Condition Is Met
For each determination zone, you can select the following items.
• Determination target waveform (see “Determination Target Waveform” described
above.)
• Enable or disable determination (ON/OFF).
• Select which condition when the waveform exits the determination zone (OUT) or
when the waveform enters the determination zone (IN) is considered true.
Determination Logic
You can select the AND logic or OR logic of the determination conditions of the six
determination zones in performing the search.
AND: Condition is considered to be true when all the determination conditions of Zone1
to Zone6 are met.
OR:
Condition is considered to be true when any of the determination conditions of
Zone1 to Zone6 is met.
Determination Operation: Sequence
You can select whether to repeat the determination operation.
Single:
Performs the determination operation once.
Continue: Repeats the determination operation until the waveform acquisition count
(described below) is reached. If the waveform acquisition count is set to
Infinite, determination operation repeats until it is aborted.
Waveform Acquisition Count: ACQ Count
You can set the number of waveform acquisitions.
Infinite:
Continues until waveform acquisition is aborted using the START/STOP key
or Abort soft key.
1-65536: Waveform acquisition and determination operation stop when the specified
number of waveforms are acquired.
10.9 Performing GO/NO-GO Determination Using Zones