10-35
IM 701730-01E
Analyzing and Searching Waveforms
10
Measuring Waveforms Using Angle Cursors (When X-Y Waveforms Are Not Displayed)
1.
Press
CURSOR
. The CURSOR menu appears.
Selecting the “Degree” Angle Cursor
2.
Press the
Type
soft key. The Type menu appears.
3.
Press the
Degree
soft key.
H&V
can also be displayed on products with firmware version 2.09 or later.
Selecting the Waveform to Measure
4.
Press the
Trace
soft key. The Trace menu appears.
5.
Press the
All
,
CH1 to CH4(2)
,
Math1
, or
Math2
soft key to select the waveform to
measure.
•
On the DL1735E/DL1740E/DL1740EL, you can select from All, CH1 to CH4, Math1, and
Math2.
•
On the DL1720E, you can select All, CH1, CH2, and Math1.
Setting the Cursor, Reference Cursor, and Reference Angle
• Moving Cursors
6.
Press the
Cursor1/Cursor2
soft key to set the jog shuttle control to Cursor1,
Cursor2, or both Cursor1 and Cursor2.
•
Select Cursor1 to move Cursor1.
•
Select Cursor2 to move Cursor2.
•
If you select both Cursor1 and Cursor2, you can move Cursor1 and Cursor2 horizontally
without changing the spacing between the two. The value of the digit being specified by
Cursor1 changes.
7.
Turn the
jog shuttle
to move the cursor.
• Moving Reference Cursors
8.
Press the
Ref1/Ref2
soft key to set the jog shuttle control to Ref1, Ref2, or both
Ref1 and Ref2.
•
Select Ref1 to move Ref1.
•
Select Ref2 to move Ref2.
•
If you select both Ref1 and Ref2, you can move Ref1 and Ref2 horizontally without
changing the spacing between the two. The value of the digit being specified by Ref1
changes.
9.
Turn the
jog shuttle
to move the reference cursor.
10.5 Cursor Measurements