Using Interleave Mode
<For a description of this function, refer to page 2-14.>
Procedure
ACTION
DELAY
X - Y
MENU
MENU
PHASE
SEARCH
HORIZONTAL
VERTICAL
TRIGGER
TRIG D
HELP
ESC
SELECT
RESET
SIMPLE
MODE
POSITION
MATH
MISC
FILE
CLEAR
TRACE
HISTORY
MEASURE
SETUP
PRESET
DISPLAY
CURSOR
CH
1
CH
2
CH
3
CH
4
POWER
COPY IMAGE SAVE
SHIFT
ZOOM
GO/NO-GO
ENHANCED
ACQ
START/STOP
SNAP
SHOT
V DIV
TIME DIV
• To exit the menu during operation, press
ESC
located above the soft keys.
• For a description of the operation using a USB
keyboard or a USB mouse, see section 4.3.
1.
Press
ACQ
. The ACQ menu appears.
2.
Press the
Interleave
soft key to select ON or OFF.
Explanation
When interleave mode is turned ON, the number of channels that can be used is limited,
but history memory, sequential store count, and record length can be set to twice the
normal values. Also, a setting of 1 GS/s becomes available in real time sampling mode.
The channels that become unavailable when interleave mode is ON are CH2 and CH4
(or just CH2 on the DL1720E).
For a description of the limitations of the record length, sample rate, and other items
when interleave mode is turned ON, see appendix 1.