6-36
IM 701210-05E
Explanation
The following four period triggers are available.
T > Time:
A trigger is activated when condition B is met the second time, if the time
when condition B is met the second time is longer than a specified time.
T < Time:
A trigger is activated when condition B is met the second time, if the time
when condition B is met the second time is shorter than a specified time.
T1 < T < T2:
A trigger is activated when condition B is met the second time, if the time
when condition B is met the second time is within a specified time range.
T < T1, T2 < T: A trigger is activated when condition B is met the second time, if the time
when condition B is met the second time is outside a specified time
range.
Setting Condition B
• Pattern of Each Channel: CH1 to CH16, Logic A, and Logic B
Select from the following:
• CH1 to CH16 (Other Than Logic Inputs)
H: Above the preset trigger level
L:
Below the preset trigger level
X:
Don’t Care
• Logic Input
Enable: Make the combination of the pattern
1
of each bit the trigger condition
Disable: Don’t Care
1. Select the pattern of each pattern from the following:
H: Above a certain level
2
L: Below a certain level
2
X: Don’t Care
2. Varies depending on the logic probe being used as follows:
700986: Approx. 1.4 V
700987: 6 V
±
50% (for DC input)
700987: 50 V
±
50% (for AC input)
• Setting the Time of Satisfaction
• For T > Time or T < Time
Time:
0.1
µ
s to 9999999.9
µ
s
• For T1 < T < T2 or T < T1, T2 < T
Time1:
0.1
µ
s to 9999999.9
µ
s
Time2:
0.2
µ
s to 10000000.0
µ
s
Setting the Trigger Level
The trigger level setting applies to both simple and enhanced triggers.
For details, see “Setting the Trigger Level” in section 6.5.
Setting the Trigger Hysteresis
Sets a width to the trigger level so that triggers are not activated by small changes in the
trigger signal. Select the trigger hysteresis from
,
, and
.
For details, see “Setting the Trigger Hysteresis” in section 6.5.
6.15 Setting the Period Trigger (ENHANCED)