
5-31
IM 701210-05E
Horizontal and Vertical Axes
5
5.17 Setting the Strain Measurement
<For a description of this function, refer to page 2-14.>
Procedure
MEASURE CURSOR
RESET
SELECT
CH
ALL CH
MODE
POSITION
SIMPLE/ENHANCED
ACQ
START/STOP
SETUP
DISPLAY
ZOOM
DUAL
CAPTURE
HISTORY
MATH
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
9
CH
10
CH
11
CH
12
CH
13
CH
14
CH
15
CH
16
DELAY
MANUAL TRIG
V/DIV
TIME/DIV
TRIGGER
TRIG D
VERTICAL
HORIZONTAL
CAL
X-Y
SEARCH
7
DSP 1
8
9
DSP 2
6
5
4
1
0
2
3
ENTER
m
DSP 6
DSP 5
LOGIC A
LOGIC B
EVENT
DSP 3
DSP 4
EXP
GO/NO-GO
ACTION
FILE
SHIFT
1.
Press a key from
CH1
to
CH16
keys to select the desired channel (select a
channel with the Strain Module (701270 (STRAIN_NDIS) or 701271
(STRAIN_DSUB)) installed).
Selecting the Measurement Range
2.
Press the
Range Unit
soft key, and then press the
µ
STR
(the unit of strain) or
mV/V
(the unit of the output value of the strain gauge transducer) soft key.
Note
Be sure to execute balancing when you change the measurement range.
• When
µ
STR Is Selected
3.
Use the
jog shuttle
to select the Range from 500
µ
STR to 20000
µ
STR.
• When mV/V Is Selected
3.
Use the
jog shuttle
to select the Range from 0.25 mV/V to 10 mV/V.
Note
If the range unit is set to mV/V, a numeric value is displayed at the right side of the range
display. This value is the maximum input at the current bridge voltage converted to volts. It
indicates the maximum input voltage at the selected range.
Setting the Display Range
4.
Press the
Upper/Lower
soft key to set the jog shuttle control to Upper.
5.
Turn the
jog shuttle
to set the upper limit.
6.
Likewise, set Lower.
Note
Pressing RESET sets the upper and lower limits to the maximum and minimum values of the
measurement range.
When Range
Unit is
µ
STR
When Range
Unit is mV/V