14. Recorder Output (Analog Output)
You can connect the lux meter to a recorder or oscilloscope and
use the recorder output function to record illuminance trends or
observe light source waveforms.
Output Signal Specifications
999 mV ± 5%
(range fixed for full scale of each range)
Load resistance: 100 kΩ or more
1.
Prepare a coated single-core shielded wire, and solder it to
the recorder accessory output plug.
Connect the other end of the wire to the input of the monitoring
instrument.
Shielded wire
Plug cover
Soldered on the – side
Soldered on the + side
2.
Attach the cap, turn on the lux meter, and insert the plug into
the recorder output connector.
[REC, R-H] appears, and the range is fixed
(manual range mode).
The automatic power-off function is disabled, and
[AUTO POWER OFF] disappears.
3.
Set the
RESPONSE
switch to FAST.
4.
1 mV of output voltage corresponds to a value of 1 in the least
significant digit of the three significant digits.
Set the appropriate range.
For details on setting the range, see chapter 6, “Range Hold”.
5.
Cover the light-detecting surface with the cap, and adjust
the zero level of the monitoring instrument.
Remove the cap and start measuring.