IM 51011-01EN
38
13.3 Detailed Description of Commands
Com
mand
51011
51012 51021
Description
HD
Yes
Yes
Sets whether to include a header in measured data sent from
the lux meter to the PC
Set whether to include a header: HDm<delimiter>
m = 0: no header, 1: with header (default)
→ Response: 0<
delimiter
>
IC
-
Yes Enables or disables the comparator during the measurement
of totalized intensity of illumination
Use the comparator: IC,m<delimiter>
m = 0: OFF, 1: ON
Note:
This applies only during the measurement of
totalized intensity of illumination.
→ Response: IC,m<
delimiter
>
m = 0: successful, 1: error
IL
-
Yes Reads the totalized intensity of illumination and time
Read the totalized intensity of illumination and time:
IL<delimiter>
→ Response: IL,mmmEsn,ttttt.t<
delimiter
>
m = mantissa of totalized value:
3 digits (000 to 999)
sn = exponent of totalized value: -2 to +3
t = integral time: 00000.0 to 10000.0 h
IS
-
Yes
Starts and stops the measurement of the totalized intensity
of illumination
Start totalized intensity of illumination: IS,m<delimiter>
m = 0: OFF, 1: ON
→ Response: IS,m<
delimiter
>
m = 0: successful, 1: error