5-1
IM 2560A-01EN
Voltage, Current, and Resistance Sourcing
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
5.1 Selecting the Voltage, Current, or Resistance
Range
Procedure
Turn the
RANGE
dial to set the voltage, current, or resistance range.
• The indicator for the selected range illuminates.
• The unit corresponding to the selected range appears.
Voltage ranges
Current
ranges
Resistance range
RANGE dial
Example: 10 V
The 10 V indicator turns on.
Explanation
Output Terminal Positions
The following figure shows the output terminal positions. The output terminals vary depending on the
selected range. For the output terminals corresponding to the selected range, see the range table
provided later.
Low
voltage-current
output terminals
(OUTPUT 1)
HI
LO
High voltage
output terminals
(OUTPUT 2)
Large current
output terminals
(OUTPUT 2)
LO HI
LO
HI
Selecting the Range
Select the range for voltage, current, and resistance.
• When the output is turned on, the signal is output from the appropriate output terminals for the
selected range.
• If you change the range when the output is on, the output turns off.
Voltage Range Settings
Select from the following.
100 mV, 1 V, 10 V, 100 V, and 1000 V
Range
Output Terminals
100 mV, 1 V, 10 V
Low voltage-current output terminals (OUTPUT 1)
The OUTPUT 1 indicator on the front panel illuminates.
100 V, 1000 V
High voltage output terminals (OUTPUT 2)
The OUTPUT 2 indicator on the front panel illuminates.
The HIGH VOLTAGE indicator will illuminate even if the output is off.
Chapter 5
Voltage, Current, and Resistance Sourcing