Cleaning and maintencance
YMC Pilot
36 / 49
PLUS
low pressure chromatography column
7.4
Replacement of spare parts
ATTENTION
Faulty assembly
Damage to the glass body possible due to faulty assembly.
► Take care during the assembly of the glass body that the maximum torque of 5
Nm for columns with 70 and 100 mm ID respectively 10 Nm for column with 140
mm and 200 mm ID are not exceeded.
► Assemble the cylinder screws in a diagonal pattern and always tighten on the
opposite sites.
► Take care during the assembly of the glass body that all parts are correctly
aligned.
7.4.1
Frit replacement in the adjustable end piece
The frit in the adjustable end piece is replaced as follows.
1. Remove the adjustable end piece Chapter 6.1, S. 18.
2. Remove the profile gasket.
3. Remove the frit.
4. Insert the new frit. The distribution profile faces the plunger.
5. Reattach the profile gasket and press firmly in position.