YL9112 Isocratic Pump
Pulsation by compressibility and delay
Compensation
of
pulsation
[Fig. 2] Delivery mechanism of YL9112 isocratic pump
2-4. Rinse port
When using buffer solution, salts are generated on back side of high pressure seal and these deposits
wear pump seal to cause shortening of seal life, which has bad effect on pump.
Rinse port enables to insert proper solvent in back side of high pressure seal to prevent salts from being
deposited and activated. Mixed solution(20% MeOH) of water and methanol is used as cleaning solution,
and life of seal is extended with lubrication action in general analysis..
6
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......