YL9110 Quaternary Pump
4-3-2. Filtration
Y
L9110
Solvent should be necessarily filtered through 0.45
μ
m or less filtering membrane before use. Removal
of small particles is necessary to compensate reliable operation of piston seal, and is necessary measure
for reliability of other components in liquid chromatograph.
Filtration process is necessary after mixing of solvent, and is more necessary in case of buffer to which
un-dissolved impurities are source of deposits. After filtration, solvent should be keep in air-tight bottle
from which small particles are removed; once solvent has been filtered, it is not necessary to filter this
solvent everyday unless reaction produce bacteria or indissoluble material occurs. If solvent is kept in
storage vessel for more than one week, it is desired to filter it again before use.
4-3-3. Solvent effect on the instrument
All parts of the Gradient Pump contacting with mobile phase is manufactured from 316 stainless steel,
ruby, sapphire, zirconium, or fluorine carbon polymer. Most of these materials are sensitive to chloride,
and it is desired to avoid use of solvent which contains even small quantity of chloride. Main solvents that
should be avoided especially are as follows.
Aqua Regia
Hydrochloric Acid(HCL) (20%)
Bromine
HCL (37%)
Chlorine Anhydrous
HCL (50%)
Copper Chloride
HCL (20%)
Ferric Chloride
HCL (75%)
Ferrous Chloride
Hydrofluorsilicic Acid (20%)
Freon 12
Hydrogen Peroxide
Guanidine
Lodine
Hydrochloride (6M)
Mercuric Chloride
Hydrobromic (20%)
(Dilute Solution)
In addition, it should be avoided to leave chloroform, carbon tetrachloride, etc. in instrument for long
time, and use of ammonium hydroxide should be avoided because it has effect on stator and rotor of
injector even though it has no effect on pump. When not using it for long time, keep it with iso-propanol
filled with in flow path.
17
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......