YL9160 PDA Detector
4-2-1. How to remove the Tungsten lamp
1) Make sure that the power cord is disconnected from the rear panel of the detector.
2) Unscrew the screws and remove the lamp assembly on the right front panel.
CAUTION
; The light of Tungsten lamp can damage eyes and skin. Always disconnect the power cord
before working in the vicinity of the lamp. The W lamp gets quite hot. Care must be taken while
handling it to prevent from burning. Always allow the lamp to cool before removing it.
3) Disconnect the Tungsten lamp from the detector by gently pulling it straight back toward you. DO
NOT twist the connector while pulling.
4) Unscrew the two thumbscrews holding the lamp mount in place, and pull the lamp mount straight
back towards you. Be careful not to lose the two thumbscrews. Be careful not to get fingerprints
on the lamp.
Y
L9160
[Fig. 6] Tungsten lamp assembly
4-2-2. How to install a new Tungsten lamp
1) Insert the new lamp assembly onto the lamp housing on the right front panel.
2) Use the thumbscrews to attach the lamp assembly to the detector.
15
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......