Using the OSICS ECL Module
OSICS User Manual
111
9.7.1
Setting Up Digital (TTL) Modulation Using the ECL Module
Internal Source
Subject
Digital modulation capabilities of the ECL module is within the 50 Hz–1 MHz frequency
range (see the ECL specifications).
Internal digital modulation uses the ECL module's built-in TTL-level signal generator (see
The ECL module’s built-in TTL signal-generator can provide a TTL modulation signal in
the range of 153 Hz–1 MHz, but for optimum modulation performance we recommend
applying modulation frequencies within the specified range of the ECL module.
The resulting modulated optical output signal follows a pattern as shown in the following
figure.
Figure 72: TTL-Modulated Output Power Signal (ECL Module)
Procedure
Step 1 – Set the module modulation source to INTERNAL:
1.
Access the ECL
Unit Setup
Accessing an OSICS Module Setup
).
2.
Expand the
Modulation
menu and enter the
Source
sub-menu.
The
Modulation source
sub-menu appears and the cursor automatically points to
the current modulation source setting.
Figure 73: ECL Unit Setup – Modulation Source (Internal)
3.
Turn the rotary knob to put the cursor before
INTERNAL
and press it to validate your
selection.
The ECL
Unit Setup
menu is automatically displayed and the modulation is set to the
selected source.
Time Reference
Power Setting
2 x Power Setting
Optical Output
Power
Main
EXIT
1
2
3
4
5
6
7
8
Unit Setup - Modulation source
ESCAPE
▲
Modulation source :
INTERNAL
MAINFRAME
Select a new value.