27
10.3
Menu
→
Measurements
→
Draft
15/06/18 10:00
DRAFT
Inlet P+
Draft 0.00 h
►Zero sensor
Save
To measure the draft:
-
Connect the manometer
hose probe to the instrument
P+
input.
-
Before starting the pressure zeroing sequence remove the gas probe from the exhaust stack.
-
Upon completion of the pressure zeroing sequence, insert the probe in the exhaust stack and
measure the draft.
INTERACTIVE OPERATION
FUNCTION
Save
The measure will be printed on the ticket of the current combustion analysis
Allows the option to print or generate the QR code for reading on the YJ Com-
bustion APP.
KEY
FUNCTION
Activate the keys shown on the bottom
lines of the display.
Returns to the previous screen.
Starts the pressure sensor autozero.
>
>
ESC
OK
Example:
15/06/18 10:00
DRAFT
5
Zero
15/06/18 10:00
DRAFT
Inlet P+
Draft --- P
►Zero
Save
15/06/18 10:00
DRAFT
Inlet P+
Draft 3.02 P
►Zero
Save
OK
B 15/06/18 10:00
WARNING
Printing
in progress ...
15/06/18 10:00
Curr. analysis
►Copy number 1
Printer BT
QR Code ON
Pairing BT