XK3190-A28E
- 8 -
after power-off. When the
【
ACCUM
】
key is pressed, if the net weight is less than 20 divisions, only
the
accumulation
value will be indicated, and the accumulation of weight and times will not be made.
6. Automatic accumulation of measured value:
Press
【
FUNCTION
】
and
【
ACCUM
】
at the same time in normal weighing state and not the peak value
holding and average value state, the indicator will enter the automatic accumulation state. In this case, the
【
ACCUM
】
light is on. In the automatic accumulation state, the indicator will perform an automatic
accumulation and indicate the times of accumulation and the results whenever the measured data is larger
than or equal to 20 divisions and after about 1-2 seconds after the data is stable. Re-press the
【
FUNCTION
】
and
【
ACCUM
】
at the same time
to exit the automatic
accumulation.
Note: automatic accumulation
state will not be saved when the power is off; however, the accumulation data will be saved.
7. Clear the accumulation result:
Press
【
CLEAR
】
in weighing state, the indicator will clear the times of accumulation and the overall
accumulation data.
8. Switching between kg and lb.:
Under the normal weighing state, the accumulation is 0 and tare weight 0,F9=3, shortly press
【
Function
】
, the
indicator performs unit switch; When the measurement unit is lb., the last digit behind decimal point of the
shown data is on. Also switch units through F8
9. Pre-set tare and upper & lower limits alarm:
Under the normal weighing status, long press [#], the steps as below
Step
Operation
Display
Description
1
[******]
Weighing indication status
2
long press[#]
[H00000]
Indicator uses to input the upper limit alarm data
3
Upper limit alarm
value, e.g. “3000”
[H03000]
Press[#] to confirm
,
enter into step “4”
4
Lower limit alarm
value, e.g. “50”
[L 00050]
Press[#] to confirm
,
enter into step “5”
5
[******]
Back to weighing status, display result after tare
【note】
cancel the upper and lower limit alarm operation when both the upper and the lower limit is 0;
Chapter 4 Calibration Instruction
Properly connect the signal source and power supply to preheat the indicator for 15-30 minutes when there is no
load on weighing platform.
After the lead sealing broken, they indicator will get into the calibration state by opening the calibration switch (it’s
approved to calibrate when leave factory, and this step can be omitted when first use of this new indicator), then
refer to the following steps (please lead sealing again after calibration):
1.Press
【
#
】
when turn on the indicator to get into the calibration state.
1
)
Set the division value
display
【
d X
】
Press
【
tare
】
to choose 1
、
2
、
5
、
10
、
20
、
50
,
press
【
#
】
to confirm,
enter into the set of next parameter automatic. Press
【
tare
】
to enter into the
cycle display automatic.
display
【
d X
】