XK3190-A28E
- 12 -
x=1:
low
x=2: medium
x=3: high
7
、
F7 x function setting
x=1:
have no other function
x=2: open average value function
x=3: open peak holding function
8
、
F8 x Kg
、
Lb setting
x=1:
Kg
x=2: Lb
9
、
F9 x auxiliary function setting
x=1:
have no auxiliary function
x=2: open internal code
x=3: open one-key change kg to Lb
Chapter 6 Error Indication
[Err 1] The signal is too small of loaded weight or the capacity of load cell is too big
[Err 2] Out of manual zero setting range
[Err 3] Zero position is too high or there is heavy object on weighing platform when starting up the
indicator.
[Err 7] The calibration short-circuit ring did not connected
[Err 8] The wrong connection of load cell signal line, please connect the load cell line in right way
[----------] Out of display range, should be within -99999
~
999999
Chapter7 Maintenance and Attention
1. To ensure the clarity and service life of the indicator, it must be kept away from direct sunlight during working,
and the ground where the indicator stands must be smooth.
2. It is improper to use this indicator in a dust or vibrant or damp environment.
3. The load cell and indicator need good connection. System must have a good ground connection, and kept away
from strong electric field, strong magnetic field. The load cell and indicator must be kept away from strong
corrosive substances and inflammable& explosive materials.
▲
!
Do not use it where inflammable gases or steams exist. Don’t use it for canning system of compressive
container.
▲
!
In the area where lightning and thunder happen frequently, reliable lightning arrester should be installed to
ensure the personal safety and to prevent any damage to the indicator and relative equipment caused by lightning
stroke.
▲
!
The load cell and indicator are both static sensitive equipment, so anti-static measures must be taken during
the use. It is strictly invalid to carry out welding operation or other operations with high current on the weighing
platform. In the stormy season, lightening prevention measures must be taken reliably to prevent any damage to
load cell and indicator caused by lightening stroke, and to guarantee the personal security of operators and safe
running of weighing devices and relative equipment.
4. Strong solvents such as benzyl and nitro oils are forbidden for cleaning the housing
5. Don’t inject any liquid or other conductive particles so as to avoid any damage of indicator and electric shock
6. Before plugging in or out of the connecting line between indicator and external equipment, the power of both
indicator and equipment should be cut off
▲
!
Before plugging in or out the connecting of load cell, it must be to cut power of indicator (power off)!
▲
!
Before plugging in or out the connecting of scoreboard, it must be to cut power of indicator and scoreboard!
▲
!
Before plugging in or out the connecting of load communication, it must be to cut power of indicator and the
upper computer!