8. Troubleshooting
Repairing procedure
Measures to be taken in case of display of water leakage,
(1) Turn OFF the power switch on the right side of main body.
(2) Dain water accumulating in the unit bottom and dry it. Remove screws fixing the sensor,
remove the electrode from the main body and wipe it well, and dry thoroughly.
(3) Upon completion of water drainage, be sure to reset the sensor to the original position.
(4) Close the door.
(5) Turn ON the power switch on the right side of main body and press the POWER key of the
operation panel. Since any faulty portion has been corrected, normal operation will begin.
Check list
Symptom
Probable cause
Power not applied
●
Faulty connection of the power cord and power cord socket outlet
●
Wire breakage of connection cord
●
Faulty power switch
Water not supplied
●
Insufficient water pressure or supply failure
●
Faulty feed water electromagnetic valve
Water supply not stopped
●
Faulty feed water electromagnetic valve
Water not sampled
●
Faulty
feed
water
electromagnetic
valve
or
sampling
electromagnetic valve
●
Clogged ion exchange resin or membrane filter
●
Faulty piping
Poor water quality
●
Degradation of CPC-P and CPC-E ion exchange resin cartridges
●
Air left in the ion exchange cartridge
●
Ion exchange resin cartridge not used over a long period of time
Leak sensor