6. Maintenance procedures
Consumables alarm display reset
Consumables alarm display reset
(
reset to be made after replacement in case of alarm
)
①
Turn ON the power switch on the left side of
main body.
②
Press the POWER key.
③
Keep pressing the CLR key of the operation
panel. The acknowledge sound is heard and
the alarm display is reset. Normal operation
can be made.
※
Be sure to perform reset after replacement of consumables.
※
In case of multiple alarms, the corresponding alarm is reset as follows. Repeat the step in the
same number of times as the number of alarms.
Reset sequence
:
CPC
→
PWF
→
FILT
※
Alarm may be output again when the water quality is not improved.
Consumables alarm display reset
(
reset to be made in case of replacement before alarm
)
①
Turn ON the power switch on the left side of
main body.
②
Press the POWER key.
③
Keep pressing the CLR key. The Water
Quality Display shows
and the
operation enters the consumables reset mode.
④
The Error
・
Consumables Replacement Display
shows
, indicating the pretreatment
cartridge. If the pretreatment cartridge is not
installed as an option,
indicating the
membrane filter is shown.
⑤
Press the
UNIT key and select the
consumables to be cleared.
⑥
Keep pressing the CLR key of the operation
panel. The acknowledge sound is heard and
the alarm display is reset. Normal operation
can be made. In order to quit the mode
halfway of operation, keep pressing the UNIT
key. The normal mode is resumed.
※
Be sure to perform reset after replacement of consumables.
※
Use this function for consumables replacement due to time-course deterioration before
appearance of consumables replacement display.
②
③
②
③
④