4. Operating procedure
Fixed temperature operation
FIXED TEMP(Fixed Temperature) mode is to keep running at target temperature. It will keep running
until operation stops.
SV
: Set Value (Target Temperature), t
:
Time
Set Fixed Temperature mode.
1 Turn on the controller.
Turn on Earth Leakage Breaker(ELB) on
(|)
the right
side wall of this Equipment. (
Idle State
)
Press and hold key to turn on the controller power.
(
Standby State
)
Indicate read temperature in Chamber on Top Screen
and indicate target temperature on Bottom Screen.
The fan motor will start and FAN lamp iluminats.
※
1
The fan motor operates when the door is closed and
stops while the door is open.
※
2
※
1,
※
2
「
The case with “FAN:S” set to “0” is excluded.
For details, refer to “Use of variable wind speed,
method of entry,” on P28,29.
2 Select Fixed Temperature Operation.
Press
key to turn FIXED TEMP (Fixed
Temperature mode) lamp on.
※
Fixed Temperature mode would be selected at first
time operation. After that, most latest operated
mode is selected.
3 Set target temperature.
①
Press
key. Then flash changeable digits on
Bottom Screen.
②
Shift to flashing digit with
key and then change
to desired digit with
keys.
Operating Temp. Range:0
~
130
℃
(Natural
convection)
0
~
270
℃
(Forced wind)
③
Press
key when target temperature setting has
completed.
Press
key once or twice to cancel its setting.
SV
t
▲
Stop operation (manually)
▲
Start operation (manually)