22
4. Operating procedures
Operating procedures (auto start operation)
This mode automatically starts fixed temperature operation after a certain time from its start set with
the timer.
However, operation does not stop automatically but needs to be stopped manually.
Procedures for auto start
operation
STO P
R U N
M EA SU R ED TEM P.
℃
H EATER
R U N
SET TEM P.
TIM ER
SU B M EN U
1. Setting an operation start time
①
After confirming the temperature you want is set, press
the TIMER key to display characters AStr
on the
MEASURED TEMP. screen that indicate auto start
operation.
The set time is displayed blinking on the set
temperature screen.
②
Set a duration you want using the
▼▲
keys.
Pressing the
▼▲
keys makes the set time blink. The
time is determined when blinking stops.
About the timer function
The maximum time that can be set for the timer is 999 hours
50 minutes.
Up to 99 hours 59 minutes, time can be set in minutes.
One hundred hours and over are set only in 10 minutes.
Keep the
▼ ▲
keys pressed to continuously change set
time and you can quickly reach the time you want.
Press
the
▼▲
keys once at a time for fine adjustment.
STO P
R U N
M EA SU R ED TEM P.
℃
H EATER
R U N
SET TEM P.
TIM ER
SU B M EN U
2. Starting timer operation
When the time you want is set, press the RUN/STOP key
while characters AStr
that indicate auto start
operation are displayed on the MEASURED TEMP. screen
and the set time on the SET TEMP. screen.
Timer starts counting when the RUN/STOP key is
pressed.
Display on the SET TEMP. screen switches from set time
display to remaining time display.
STO P
R U N
M EA SU R ED TEM P.
℃
H EATER
R U N
SET TEM P.
TIM ER
SU B M EN U
3. Stopping and ending timer operation
Operation automatically starts at the set time and the RUN
lamp comes on.
To stop operation, press the RUN/STOP key for approx.
one second to end the timer operation mode. The screen
switches to the initial setting screen.
①
②
Summary of Contents for DKM300C