21
8. TROUBLESHOOTING
Troubleshooting Guide
Symptom
Possible Causes/Solutions
Unit will not power on.
Power cable is not connected properly or securely.
Power failure in progress.
Main power switch (ELB) is
“OFF”
Temperature does not fall.
Cooling fins are clogged with dust or debris.
Too many samples are being processed at once.
External temperature exceeds 30°C.
Cooling vent(s) are obstructed.
Error lamps
Error symptom
Cause
Remedy
Refrigeration
error lamp on
Refrigeration system
overload
Turn power off immediately, eliminate error source
(refer to "Trouble Shooting"
– “temperature does not
fall
”, above) and restart after 5 minutes.
If lamp stays lit after restart, call for service.
If problem persists, turn off power immediately, disconnect power cable and call for service.
Summary of Contents for DC401